The "no breakthrough alarm if oxygen fails problem".

Silencing the alarm for any reason will also disable the oxygen supply failure alarm. This does not have a break-through feature and if there is subsequent oxygen supply failure within the next 2 minutes there will be no audible alarm and the patient will not be ventilated.

The Oxylog® 3000 relies on an oxygen supply to drive the ventilator.