Draeger primus
Incorrect seating of the water trap adapter, if installed, (not just the seating of the water trap in the adapter) may cause a circuit leak. refl.in/co
Improve to correct grammar or spelling, but do not significantly alter the substance of the original author's text.
Cancel
Endorsed by
,
Test account
,
more...
Private
only
only
Apr 21, 2013
It is possible for the CO2 tubing to become caught in the APL valve. It may cause altered or absent capnography trace. refl.in/lv
At our institution we feed the CO2 tubing through the handle to avoid any potential problems.
At our institution we feed the CO2 tubing through the handle to avoid any potential problems.
Improve to correct grammar or spelling, but do not significantly alter the substance of the original author's text.
Cancel
Endorsed by
,
Pieter Peach
,
more...
Private
only
only
Sep 20, 2012
In the event of ventilator failure, pushing in and popping out the backup fresh gas flow dial delivers gas independently of the ventilator via the back bar (so it passes through the volatile)
Improve to correct grammar or spelling, but do not significantly alter the substance of the original author's text.
Cancel
Endorsed by
,
more...
Private
only
only
Apr 17, 2013
The machine can be placed in a "monitoring only" mode (no fresh gas flow) by pressing either:
1. Monitoring mode button
2. Press power button to the right of the wheel when in standby mode.
1. Monitoring mode button
2. Press power button to the right of the wheel when in standby mode.
Improve to correct grammar or spelling, but do not significantly alter the substance of the original author's text.
Cancel
Endorsed by
,
more...
Private
only
only
Apr 17, 2013